      The HT R&D laboratory having Hydro Turbine/Pump Model Testing test bed and Water Flow calibration Facility, established in 2018, has been providing technical consultancy, research, training, testing and calibration in flow measurement for our clients across the globe. Activities of Water Flow calibration Facility are centered around calibration and testing of flow products including flow meters, valves with water as medium of flow.

 

Prominent Features

  • Gravimetric Calibration method as per ISO 4185:1980
  • Comparison Calibration method using reference flow meters
  • Accredited as per ISO/IEC 17025:2017 accredited by National Accreditation Board for Testing and Calibration Laboratories (NABL)
  • Full compliance and audit readiness for your flow meter with complete and traceable calibration
  • Secure accurate and repeatable results for your devices over their entire life cycle by putting our expertise
  • The software is written in LabVIEW written by HT R&D lab as this has proved to be the most effective platform for calibration excellence and gives freedom to develop and introduce improvements at any time along with Software validation
  • Temperature control during calibration by provision of chiller plant during calibration
  • Ensuring constant flow-rate by use of transfer pumps to maintain inlet water level during calibration

Different size Flow-​meter Calibration Loop

Flow Diverter

Specifications of the Water Flow calibration Facility

  Flow Rate Range     Flow Velocity Range     Uncertainty in Volume Flow Rate  
75 Lit/S to 907 Lit/s 0.8 m/s to 10 m/s ±0.18% (ISO 4185)

Top...

 

Salient Technical Details

  • Measuring tank large capacity to collect 45 tons of water, on weighing system of 3 load cells 22 tons each
  • Measuring tank calibration by calibration tank with 2 ton load cell
  • Moving Flap Flow Diveter for higher accuracy and better control
  • Accurate diversion time measurement by 12 digit precision counter triggered by Photo switch
  • Flow calibration range 75 l/s to 907 l/s
  • Flow velocity calibration range 0.8m/s to 10 m/s
  • Two pumps driven by VFD motors in parallel/series or solo mode for stable flow in full range
  • Facility for inline installation of flow-meter from 100-350 NB

 

 
       
    Contact:

Hydraulic Turbine R&D Laboratory,
Department of Hydro and Renewable Energy,
Indian Institute of Technology Roorkee
Roorkee-247667, Uttarakhand, India

Phone : Off.(+91 1332) 286133, 286134, 285213
E-mail :hturbinelab@iitr.ac.in,
hturbinelab@gmail.com,
arun.kumar@hre.iitr.ac.in