UV-VIS Spectrophotometer
The UV-1800 is an advanced high-resolution spectrophotometer utilizing a precision Czerny-Turner optical system. Operation can be either as a stand-alone instrument or as a PC-controlled instrument with the included UV Probe software.
About Instrument and its Applications:
The UV-1800 is an advanced high-resolution spectrophotometer utilizing a precision Czerny-Turner optical system. Operation can be either as a stand-alone instrument or as a PC-controlled instrument with the included UV Probe software. USB memory can be connected directly to the UV-1800 for simple data transfer and printing is possible with the optional screen-copy printer or printers that support PCL control codes.
Photometric Mode measures the absorbance or transmittance at a single wavelength or at multiple (up to eight) wavelengths. In multiple-wavelength measurement, calculations can be performed on the data obtained for up to four wavelengths, including the calculation of the difference between, or ratio of, the measurements obtained for two wavelengths.
Spectrum Mode obtains sample spectra using wavelength scanning. Changes in the sample can be tracked using repeated scans. Data processing operations, such as the enlargement or reduction of the spectra obtained, peak detection, and area calculations, are also possible.
Quantitation Mode generates a calibration curve from standard samples, and uses it to calculate the concentrations of unknown samples. Various combinations of wavelengths (1 to 3 wavelengths and derivatives) and calibration curves (K-factor and first-to-third order regression calibration curves) are possible.
Kinetics Mode measures the change in absorbance as a function of time, and calculates enzymatic activity values. You can select either the kinetics measurement method or the rate measurement method.
Time-Scan Mode measures the change in absorbance, transmittance, or energy as a function of time.
General Information-
Make - Shimadzu
Model – UV 1800
Year of Installation - 2013
Specifications-
Stray Light: <0.02%T at 340 nm, 400 nm; <1.0%T at 198 nm
Source Lamp Deuterium; Tungsten-Halogen
Source Lamp Life 2000 hours
Beam Type Double Bandwidth 1.0 nm Min Wavelength (nm)190 Max Wavelength (nm)1100
Wavelength Accuracy ±0.1 nm
Wavelength Reproducibility ±0.1 nm Min
Photometric - Transmittance (%T)0
Max Photometric - Transmittance (%T)400
Min Photometric - Absorbance (A)-4 Max Photometric - Absorbance (A)4
Photometric Drift <0.0003 A/hr
Detector Silicon photodiode
Output USB
Display Type: LCD (4.75" x 3.5")
Sample Requirements-
Liquid and Film type Samples.